15510-EN Rev G Troubleshooting 25
Problem User Action(s)
The treatment lesions are variable
or intermittent (LIO only)
The LIO may be slightly out of focus. This decreases power density.
Readjust your working distance to obtain the smallest spot size.
A poorly centered laser beam may be clipping on the examination lens or
on the patient’s iris. Adjust the laser beam in the illumination eld.
The laser treatment parameters may be too close to the tissue response
threshold for consistent response. Increase the laser power and/or
exposure duration, or select a different lens.
Does not t on the mounting plate
(OMA only)
Inspect and clean the mounting plate.
Verify that the mounting plate corresponds to your microscope.
Laser and viewing systems are
not focussed at the same point
(OMA* only)
Verify installation of a 175 mm microscope objective lens on
the microscope.
Turn on the aiming beam to determine focus position and adjust
as necessary.
View is blocked or partially
blocked by OMA (OMA* only)
Set magnication to 10X or more.
* Operating microscope adapter compatible with IRIDEX IQ 810 and SLx Systems.