15510-EN Rev G Maintenance 31
Power Measurements using a CW Delivery Device
Power (mW) Exposure Duration (ms) Meter Reading (mW) Acceptable Range (mW)
200 1000–3000 160–240
500 1000–3000 400-600
1000 1000–3000 800-1200
2000 1000–3000 1600-2400
Data for power measurement equipment: _______________ Calibration date: _______________
Meter Model and Serial Number: _______________ Calibrated by: _______________
Power Measurements using a MicroPulse
®
Delivery Device
Exposure
Duration (ms)
MicroPulse
®
Duration (ms)
MicroPulse
®
Interval (ms)
Indicated Power
(mW)
Measured
Power (mW)
Acceptable
Range (mW)
1000–3000 1.0 1.0 200 80-120
1000–3000 1.0 1.0 500 200-300
1000–3000 1.0 1.0 1000 400-600
1000–3000 1.0 1.0 2000 800-1200
Data for power measurement equipment: _______________ Calibration date: _______________
Meter Model and Serial Number: _______________ Calibrated by: _______________