EasyManua.ls Logo

ISC Tronic - Operation Parameter Setting Menu

ISC Tronic
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rev. 0128/01
TRONIC
SERVICE MANUAL
SYSTEMS
6
1.5) OPERATION PARAMETER SETTING MENU
WARNING: PERFORM THIS PROCEDURE ONLY AFTER HAVING READ THE
FOLLOWING INSTRUCTIONS CAREFULLY. IF NOT PERFORMED PROPERLY, THIS
PROCEDURE CAN CAUSE MALFUNCTIONING AND DAMAGE THE UNIT.
Set menu 0567 (see chap. 1.1),confirm with the S Key.
Press the F key repeatedly to select the following operation parameters in sequence:
CODE
VALUE
P-t02
60
P-t03
180
P-t04
60
P-t05
45
P-t06
04
P-t07
05
P-t08
02
P-t09
60
P-t10
00
P-t11
09
P-t12
60
P-t13
05
P-t14
240
P-t15
05
P-udL
0.700
P-udH
1.400
P-ud1
0.120
P--Fd
1.000
rES
P--Lb
10.000
t--AL
3600
When the display shows u---F, the pressure transducer can be calibrated.
Disconnect the quick connectors from the high and low pressure hoses.
Open the LP-HP valves on the control panel to connect the pressure transducer to the
atmospheric pressure.
With the display showing u---F, confirm with the S key.
The display shows the theorical weight measured.
Digit the value 0.960. Confirm with the S key.