TRONIC
SERVICE MANUAL
SYSTEMS
6
1.5) OPERATION PARAMETER SETTING MENU
WARNING: PERFORM THIS PROCEDURE ONLY AFTER HAVING READ THE
FOLLOWING INSTRUCTIONS CAREFULLY. IF NOT PERFORMED PROPERLY, THIS
PROCEDURE CAN CAUSE MALFUNCTIONING AND DAMAGE THE UNIT.
Set menu 0567 (see chap. 1.1),confirm with the S Key.
Press the F key repeatedly to select the following operation parameters in sequence:
When the display shows u---F, the pressure transducer can be calibrated.
Disconnect the quick connectors from the high and low pressure hoses.
Open the LP-HP valves on the control panel to connect the pressure transducer to the
atmospheric pressure.
With the display showing u---F, confirm with the S key.
The display shows the theorical weight measured.
Digit the value 0.960. Confirm with the S key.