EasyManua.ls Logo

ITech IT6821 - Output Order

ITech IT6821
14 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
Parameter:0~255
Reset value:Consult *PSC order
Example:STATus:QUEStionable:ENABle 128
Quest syntax:STATus:QUEStionable:ENABle?
Return parameter:<NR1>
Reference order:*PSC
STATus:OPERation:EVENt]?
This order can read the parameter from operation event register. After executing this order,
operation event register is reset.
Quest syntax:STATus: OPERation [:EVENt]?
Parameter:None
Return parameter:<NR1>
Reference order: STATus: OPERation:ENABle
Bit determination of operation event enable register
Bit Position 7 6 5 4 3 2 1 0
Bit Name no use no use no use no use no use no use CC CV
Bit value 2 1
STATus:OPERation:CONDition?
This order can read the parameter from the operation condition. When the parameter of
operation condition register changes, the bit corresponding in operation event register is 1.
Quest syntax:STATus: OPERation: CONDition?
Parameter:None
Return parameter:<NR1>
STATus:OPERation:ENABle
This order can set the parameter of operation even enable register. Setting parameter can
determine which bit value of operation event register is 1 and the bit will enable OPER of
status byte register is 1.
Order syntax:STATus: OPERation:ENABle <NRf>
Parameter:0~255
Reset value:Consult *PSC order
Example:STATus: OPERation:ENABle 128
Quest syntax:STATus: OPERation:ENABle?
Return parameter:<NR1>
Reference order:*PSC
4.3 Output Order
ONPut[:STATe]
This order can set power supply output on or off..
Order syntax:ONPut[:STATe] <bool>
Parameter:0|1|ON|OFF
*RST value:OFF