14
MEASure[:SCALar]:CURRent[:DC]?
This order can get the input current of power supply.
Order syntaxMEASure[:SCALar]:CURRent[:DC]?
ParameterNone
Return parameterNR2
Return parameter unitA
ExampleMEAS:CURR
MEASure[:SCALar]:POWer[:DC]?
This order can get the input power of the power supply.
Order syntaxMEASure[:SCALar]:POWer?
ParameterNone
Return parameterNR2
Return parameter unitW
ExampleMEAS:POW?
Calibration order
CALibration:SECure:[STATe]
Set protection mode enable or disable when calibrating the power supply.
Order syntaxCALibration:SECure:[STATe ]{ON|OFF>,[<password>]}
Parameter0|1|ON|OFF, ‘5811
ExampleCAL:SEC 1, ‘5811; CAL:SEC OFF
Quest syntaxCALibration:SECure:STATe
ParameterNone
CALibration:VOLTage:LEVel
This order can set voltage calibration point. P1P2P3P4 must be calibrated orderly.
Order syntaxCALibration:VOLTage:LEVel <point>
ParameterP1|P2
CALibration:VOLTage [:DATA] {<numeric value>}
Return actual output voltage value of calibration point.
Order syntaxCALibration:VOLTage [:DATA] <NRf>
Parameter<NRf>
ExampleCAL:VOLT 30.0002V
CALibration:CURRent:LEVel
This order can set current calibration point. P1P2P3P4 must be calibrated orderly.
Order syntaxCALibration:CURRent:LEVel <point>
ParameterP1|P2
CALibration:CURRent [:DATA] {<numeric value>}
Return actual output current value to calibration point.
Order syntaxCALibration:CURRent [:DATA] <NRf>
Parameter<NRf>
ExampleCAL:VOLT 3.0002A