14
MEASure[:SCALar]:CURRent[:DC]?
This order can get the input current of power supply.
Order syntax:MEASure[:SCALar]:CURRent[:DC]?
Parameter:None
Return parameter:〈NR2〉
Return parameter unit:A
Example:MEAS:CURR?
MEASure[:SCALar]:POWer[:DC]?
This order can get the input power of the power supply.
Order syntax:MEASure[:SCALar]:POWer?
Parameter:None
Return parameter:〈NR2〉
Return parameter unit:W
Example:MEAS:POW?
Calibration order
CALibration:SECure:[STATe]
Set protection mode enable or disable when calibrating the power supply.
Order syntax:CALibration:SECure:[STATe ]{ON|OFF>,[<password>]}
Parameter:0|1|ON|OFF, ‘5811
Example:CAL:SEC 1, ‘5811; CAL:SEC OFF
Quest syntax:CALibration:SECure:STATe?
Parameter:None
CALibration:VOLTage:LEVel
This order can set voltage calibration point. P1、P2、P3、P4 must be calibrated orderly.
Order syntax:CALibration:VOLTage:LEVel <point>
Parameter:P1|P2
CALibration:VOLTage [:DATA] {<numeric value>}
Return actual output voltage value of calibration point.
Order syntax:CALibration:VOLTage [:DATA] <NRf>
Parameter:<NRf>
Example:CAL:VOLT 30.0002V
CALibration:CURRent:LEVel
This order can set current calibration point. P1、P2、P3、P4 must be calibrated orderly.
Order syntax:CALibration:CURRent:LEVel <point>
Parameter:P1|P2
CALibration:CURRent [:DATA] {<numeric value>}
Return actual output current value to calibration point.
Order syntax:CALibration:CURRent [:DATA] <NRf>
Parameter:<NRf>
Example:CAL:VOLT 3.0002A