14 = Returning (agitate mode)
18 = Agitate Dwelling
26 = A fault is being cleared
r => DISPENSE RATE
r or r0 r0,<value2> Returns the current Production mode rate.
r0,<value2> r0,<value2> Sets the current Production mode rate.
<value2>: represents a 4 digit decimal number
Maximum: 1500 (dependent on Motor/Base selection)
Minimum: 1
Default: 150
r1 r1,<value2> Returns the current Fluidic mode.
r1,<value2> r1,<value2> Sets the current Fluidic mode.
<value2>: represents a 4 digit decimal number
Maximum: 1500 (dependent on Motor/Base selection)
Minimum: 1
Default: 600
s => STATUS
s0 s0,<value2> Returns the volume remaining in the pump during the current operation (strokes)
s2 s2,<value2> most recent number of rotary (spindle) stalls as <value2>.
Maximum: 15
Minimum: 0
s3 s3,<value2> Dump parameters, multiple line response.
s4 s4,<value2> Dump status, multiple line response.
s5 s5,<value2> Returns the piston location relative to the 0 degree Stop Position (millirevs).
s8 s8,<value2> Returns the current Fault Code.
s9 s9,<value2> Returns the current Error Code.
s10 s10,<value2> Returns the Assert SW Number.
s11 s11,<value2> Returns the Assert SW Code.
s12 s12,<value2> Returns the Assert HW Number.
s13 s13,<value2> Returns the Assert HW Code.
s14 s14,<value2> Returns the Serial Number.
s15 s15,<value2> Returns all parameters contained in a recipe.
s16 s16,<value2> Returns all parameters not contained in a recipe
s20 s20,<value2> Returns the present state of the LOGIC I/O and Auxiliary Outputs.
t => DWELL
t2 t2,<value2> Returns the current Pre-Op Dwell in seconds.
t2,<value2> t2,<value2> Sets the Pre-Op Dwell.
<value2>:
Maximum: 60.00
Minimum: 0.00 (default)
t3 t3,<value2> Returns the current Post-Op Dwell in seconds.
t3,<value2> t3,<value2> Sets the Post-Op dwell.
<value2>:
Maximum: 60.00
Minimum: 0.00 (default)
t4 t4,<value2> Returns the current Autotrigger Dwell in seconds.
t4,<value2> t4,<value2> Sets the Autotrigger Dwell.
<value2>:
Maximum: 300.00
Minimum: 0.00 (default)
t5 t5,<value2> Returns the Agitate Dwell in seconds.
t5,<value2> t5,<value2> Sets the Agitate Dwell .
<value2>:
Maximum: 300.00
Minimum: 0.00