Behaviour of the furnace in case of error
* Furnace head opens when this error occurs.
** A program in progress is stopped.
*** The error cannot be acknowledged; the programs cannot be started.
Index
Category
Error ERR Conti- Error Message Text
No. nuation
possible
57 System Technical error 1024 no Error while reading the stop switch for the furnace head.
of the furnace head **,***
58 System Technical error of the 1025 no Read/write CPLD
furnace head
**, ***
59 System Technical error of the 1026 no Error in the vacuum driver
vacuum driver
60 System Technical error in the 1028 Error while reading/writing the SRAM.
SBS driver
67 System Reading, calculating the 1202 no Error while measuring the ambient temperature
ambient temperature
**,***
68 System Reading, calculating the 1203 no Error while measuring the furnace temperature
furnace temperature
**,***
69 System Reading, calculating the 1204 no Error while measuring the furnace control temperature
furnace control temperature **,***
70 System Reading, caluclating the 1205 Error while measuring the resistance value for the ATK2 calibration.
resistance value
71 System Reading, calculating, 1206 Error while measuring the temperature for the EXTERNAL T-sensor.
EXTERNAL T-sensor
72 System Temperature regulator 1207 no Error in the temperature regulator.
**,***
80 System Access Board Descriptor: 1400 Error while writing the new version number to the BoardDescriptor (E2Prom).
Writing the version number
81 System Access Board Descriptor: 1401 Error while writing the new serial number to the BoardDescriptor (E2Prom).
Writing the serial number
82 System Access Board Descriptor: 1402 Error while reading the serial number from the BoardDescriptor (E2Prom).
Reading the serial number
83 System
Loading of the drivers failed
1500 Failure during loading the necessary drivers. The furnace is not ready. Contact your Ivoclar Vivadent Service
*** Center.