9. Product Specifications
9.1 Delivery form
Programat P300
Power cord
Vacuum hose
Calibration Test Pack
Operating Instructions
Programat Firing Tray Kit
9.1.2 Recommended accessories
Programat Accessories Set
Temperature Checking Set 2
Vacuum Pump VP3 / VP3 easy
9.2 Technical data
Power supply 110–120 V / 50–60 Hz
200–240 V / 50–60 Hz
Overvoltage category II
Contamination level 2
Tolerated voltage fluctuations +/- 10%
Max. power consumption 12 A at 110–120 V
8.5 A at 200–240 V
Acceptable data for vacuum pump of other manufacturers
Max. output: 250 W / max. leakage current 0.75 mA
Final vacuum: < 50 mbar
Use only tested pumps
Electrical fuses: 110–120 V:
250 V / T 15 A (heating circuit)
250 V / T 5 A (vacuum pump)
200–240 V:
250 V / T 8 A ( heating circuit)
250 V / T 3.15 A ( vacuum pump)
Dimensions of electrical fuses 110–120 V:
Diameter 6.3 x 32 mm
200–240 V:
Diameter 5 x 20 mm
Dimensions of the closed furnace
Depth: 368 mm / width: 303 mm / 398 mm (with Cooling tray)
Height: 320 mm
Usable size of the firing chamber Diameter 80 mm
Height 48 mm
Max. firing temperature 1200 °C (2192 °F)
Weight Furnace base: 12.3 kg
Furnace head: 4.5 kg
Safety information
The P300 complies with the following guidelines:
IEC 1010-1/EN 61010, Part 1
– UL and cUL standards
Radio protection / electromagnetic compatibility EMC tested
9.3 Acceptable operating conditions
Acceptable ambient temperature range:
+5 °C to +40 °C (+41 °F to +104 °F)
Acceptable humidity range:
80 % maximum relative humidity for temperatures up to 31 °C
(87.8 ° F) gradually decreasing to 50 % relative humidity at 40 °C
(104 °F); condensation excluded.
Acceptable ambient pressure:
The furnace is tested for use at altitudes of up to 2000 m above sea
level.
9.4 Acceptable transportation and storage conditions
Acceptable temperature range -20 to +65 °C (-4 °F to +149 °F)
Acceptable humidity range Max. 80 % relative humidity
Acceptable ambient pressure 500 mbar to 1060 mbar
Use only original packaging of the Programat P300 together with the