24 System Checksum error in the 45 Invalid checksum of the memory for firing program data - the firing program data are written to the internal
firing program memory memory using the original values.
25 System Write error in the 46 Error while writing firing group data to the internal memory.
firing group memory
26 System Read error in the 47 Error while reading the firing group data from the internal memory.
firing group memory
27 System Checksum error in the 48 Invalid checksum of the memory for firing group data - the firing group data are written to the internal
firing group memory memory using the original values.
28 System OT1 <> OT2 54 nein Error in the temperature measuring circuit! Contact your local Ivoclar Vivadent Service Center.
plausibility check:
**,***
OT1 = OT2 +/- 10°C
29 System Temperature in the furnace 56 Temperature in the furnace base is lower than 1 °C. Bring the furnace base to a higher operating temperature.
base is lower than 1 °C
30 System Program start blocked 103 Starting a program is not possible due to a technical malfunction.
31 System Incorrect time settings 107 The clock settings are incorrect. Please set a correct date and a correct time!
(date / time)
33 Entry HV > H (H2) 110 Enter a lower value for HV or a higher value for H (H2)
35 Entry "Share of the holding time 120 Activate the vacuum during the holding time Tx or deactivate HV..
with vacuum" is activated,
but Vx2 does not
correspond to Tx or Tx+1
36 System Supply voltage outside the 700 The supply voltage is outside the acceptable range. Check the supply voltage.
acceptable range
37 System Start-up aborted due 701 The self-test of the furnace was interrupted by an error. It is not possible to work with the furnace! Switch
to an error *** the furnace off and on again, once the error has been rectified.
38 System Brief power failure during a 702 A firing program in progress was interrupted by a brief power failure. The program is continued!
firing program in progress
39 System Power failure during a 703 A firing program in progress (started from the USB memory stick) was interrupted by a power failure.
firing program from the The program could not be continued, since the USB memory stick is no longer present!
memory stick in progress –
memory stick no longer
present.
40 System Prolonged power failure 704 An overnight program (firing program) in progress was interrupted by a prolonged power failure.
during an onvernight The overnight program is continued!
program in progress
41 System Reading and processing 705 no Error during measuring the supply voltage. Contact your local Ivoclar Vivadent Service Center.
supply voltage
**,***
42 System
Reading the power frequency
706 Error during measuring the supply voltage. Contact your local Ivoclar Vivadent Service Center.
43 System Incorrect supply voltage 707 The furnace is operated with the incorrect supply voltage. Make sure that the furnace is operated with the
supply voltage indicated on the rating plate.
44 System Final vacuum value not 800 The required final vacuum value cannot be reached. Check the vacuum pump.
reached
45 System Vacuum drop 801 An unacceptable vacuum drop has occurred.
46 System The vacuum does not 802 No vacuum increase could be measured. Check the following points: Is the firing chamber tight (no
increase (self-test) contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?
Is the fuse F1 o.k.?
47 System Temp. EXTERNAL T-SENSOR 1010 Temperature channel EXTERNAL T-sensor excess temperature
excess temperature
(> 1225°C)
48 System Write error in the furnace 1011 Error while saving furnace configuration data to the internal memory.
configuration data memory
49 System Read error in the furnace 1012 Error while reading the furnace configuration data from the internal memory.
configuration data memory
50 System Checksum error in the 1013 Invalid checksum of the memory - furnace configuration data are written into the internal memory using
furnace configuration the original values.
data memory
51 System Write error in the furnace 1014 Error while saving the furnace operational data to the internal memory.
operational data memory
52 System Read error in the furnace 1015 Error while reading the furnace operational data from the internal memory.
operational data memory
53 System Checksum error in the 1016 Invalid checksum of the memory – furnace operational data are written into the internal memory using
furnace operational data the original values.