RAC Plus User’s Manual
5-4
A-1
Q. My RAC will not turn on. What’s wrong?
A. There are several possible causes for this.
1. You may have a loose connection. Double check all connections
at the MDS and RAC to make sure they are tight and at the correct
locations.
2. There may be a problem with the telephone jack on the power/
signal cable. Check to make sure the contacts are not bent.
3. Check the fuse between the MDS and the battery. If the internal
lament is broken the fuse has been blown.
Q. My RAC counts in increments of 2, 5, 10, etc. What’s wrong?
A. Your calibration number is too high. The calibration number should be
between .500 and 1.200. You can lower the calibration number by adjusting
the rotary switch on your modular distance sensor. Refer to Chapter 2 for
more detailed information.