Remote Operation Getting Started Guide
Example remote control procedures
24 Transport Module, DMC, MSAM, and 5800 Remote Operation
> :SENSe:DATA? CSTatus:PDH:DS1:AIS
Query the AIS alarm result
1
If the unit is self-looped, it will detect the AIS.
> :SOURce:PDH:DS1:INSert:AIS OFF
Turn the alarm back off
> :SENSe:DATA? CSTatus:PDH:DS1:AIS
Query the AIS alarm result again
0
Configuring a timed
test
There are three test timing modes: not timed, timed test, and delayed-start timed test. This
example demonstrates how to configure each of those.
> :SENSe:TEST:ENABle OFF
Sets the test type to "Not Timed"
> :SENSe:TEST:ENABle ON
Sets the test type to "Timed Test"
> :SENSe:TEST:DURation 3600
Set the timed test duration in seconds
> :SENSe:TEST:ENABle DELAYED
Sets the test type to "Delayed-start timed test"
> :SENSe:TEST:DURation 3600
Set the timed test duration in seconds
> :SENSe:TEST:DELay:DATE "10/31/2011"
Set the start test date
> :SENSe:TEST:DELay:TIME "23:00:00"
Set the start test time
Exiting an application
To shut down an application through remote control, it must first be selected. If the applica-
tion is already selected and a session has been created, that session should be ended.
> :SYSTem:APPLication:SELect TermDs1Bert_121
> :EXIT
or
> :SESSion:END
> :EXIT