Do you have a question about the Jeio tech OV-12 and is the answer not in the manual?
Covers symptoms like temperature not increasing, trembling, or rising, and their related malfunctions and checks.
Troubleshooting steps for when the unit does not power on, including fuse and switch checks.
Diagnosing and resolving power interruptions that occur while the unit is running, including PCB and over current issues.
Addresses power loss immediately after the unit starts, focusing on electric capacity and heater leakage.
Troubleshooting when the circuit breaker trips, investigating heater leakage and wire insulation errors.
Measuring the output voltage of the relay on the main control board after pressing the start key.
Checking the heater's output voltage by setting temperature and observing the HEAT LED.
Measuring the AC input voltage of the main control board after switching the unit ON.
Measuring the resistance of the heater to detect short circuits.
Measuring the resistance of the heater to detect current leakage.
| Model | OV-12 |
|---|---|
| Category | Laboratory Equipment |
| Type | Oven |
| Temperature Range | Ambient +5°C to 250°C |
| Capacity | 32 liters |
| Shelves | 2 |
| Controller | Digital PID controller |
| Timer | 99 hr 59 min |
| Material (Interior) | Stainless steel |
| Material (Exterior) | Powder coated steel |
| Power Supply | 220V, 50/60Hz |
| Temperature Accuracy | ±1°C at 100°C |