ERROR DISPLAYS
Main
Display
Secondary
Display
Mode/State Reasons
OuEr
OuEr
-10.0~120.0℃
OuE/Und
Uncompensated
conductivity
measurement
Uncompensated conductivity is out of the range.
Temperature is out of the -10.0~120℃ range.
OuEr
OuEr
LErr
-10.0~120.0℃
OuE/Und
-10.0~120.0℃
Compensated
conductivity
measurement
Compensated conductivity is out of the range.
Temperature is out of the -10.0~120℃ range.
Temperature exceeds the limit (compute using the current temperature
coefficient and temperature reference).
OuEr
OuEr
-10.0~120.0℃
OuE/Und
Salinity
measurement
Salinity value is out of the range.
Temperature is out of the -10.0~120℃ range.
OuEr
OuEr
LErr
-10.0~120.0℃
OuE/Und
-10.0~120.0℃
TDS
measurement
TDS value is out of the range.
Temperature is out of the -10.0~120℃ range.
Temperature exceeds the limit (Compute using the current temperature
coefficient and temperature reference).
OuEr
Undr
Cell constant
calibration
Cell constant calibration is greater than ±30% of the selected basic cell
constant.
OuE/Und Temperature
measurement
Temperature is out of the -10.0~120℃ range.