EasyManua.ls Logo

JENCO 3173 - Calibration Procedures; Calibration Process Steps

JENCO 3173
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
To change between modes, simply press and release the [MODE] key, the unit will cycle
through these four modes. When change mode, please carefully observe the annunciators
on the LCD.
Note : Every time the unit is turned OFF the last mode will be saved so that when turn the
instrument ON again it will return to this mode.
CALIBRATION SET-UP
BEFORE CALIBRATE
System calibration is rarely required as Model 3173/3173R were all factory pre-calibrated.
However, from time to time it is wise to check the system calibration and make
adjustment when necessary.
When the new probe / instrument is used, CALIBRATION should be performed, since
the cell constant of conductivity probes may differ from each other.
CALIBRATION PROCESS
1. After the power-on diagnostics, press the [CAL] key and the CAL annunciator will
appear on the display.
TEMPERATURE COEFFICIENT
2. The unit will display 1.91% or a value set previously using
this procedure. Use the [ ] and [ ] keys to change the
value to the desired new temperature coefficient.
3. Press the [ENTER] key to save the new value of
temperature coefficient or press the [MODE] key to exit
(and go to the next calibration parameter).
TEMPERATURE REFERENCE
4. The unit will display 25.0C or a value set previously using
this procedure. Use the [ ] and [ ] keys to change the
value to the desired new temperature reference.
5. Press the [ENTER] key to save the new value of

Related product manuals