EasyManua.ls Logo

JENCO 9031M - Specifications

JENCO 9031M
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17
SPECIFICATIONS
Display Range Resolution Accurac y
Dissolved Oxygen 0.00 to 20.00
mg/L
0.01 mg/L
±1.5% of
reading or
±0.15 mg/L
whichever is
greater
Dissolved Oxygen
% saturation
0.0 to 200.0% 0.1%
±1.5% of
reading or
±1.5% air
saturation,
whichever is
greater
Temperature 0.0 to 50.0 °C 0.1 °C ±0.3 °C
Pressure compensation 500 to 1125 mBar
Salinity compensation
0.0 to 40.0 ppt
Calibration Up to 2 point
Calibration Back-up EEPROM
Datalogging capabilities
50 data sets
Automatic shut off function 10 minutes of non-use
Audio Feedback All Touch Keys
Power
9 Volt battery
Battery Life
50 hour battery life in
manual
Ambient Temperature Range 0 to 50 °C
Relative Humidity
At 90% RH
Case IP67
waterproof
Dimensions
214 x 74 x 42 mm
216 X 80 X 50 mm (Have
jacket)
Weight 290 grams
* Automatic dissolved oxygen temperature compensation range is 0
to 45 °C