EasyManua.ls Logo

Julabo MA-4 - Temperature Application to External Systems

Julabo MA-4
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MA
Caution:
Fire or other dangers when using bath fluids that are not recommended:
Please contact JULABO before using other than recommended bath fluids.
Use only nonacidic and noncorrosive bath fluids.
JULABO assumes no liability for damage caused by the selection of an
unsuitable bath liquid.
Unsuitable bath fluids are fluids which, e.g.,
• are highly viscous
(much higher than recommended at the respective working temperature)
• have a low viscosity and have creep characteristics
• have corrosive characteristics or
• tend to crack.
• No liability for use of other bath fluids!
5.3. Temperature application to external systems
Caution: Securely attach all tubing to prevent slipping.
If the circulator is operated without external system, close the pump connector
(24a) with the cap nut.
feed return
The circulator is used for temperature application to
external, closed systems (loop circuit) with simultaneous
temperature application in the circulator bath.
Connecting the external system
• Unscrew the collar nuts from the pump connector (24a).
• Slide the tubing onto the pump connector for feed (24a)
and return flow (24b) and secure with hose clamps.
17