EasyManua.ls Logo

Julabo MA-4 - Page 47

Julabo MA-4
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MA
ERROR MESSAGES
Description
-06 SENSOR DIFFERENCE ALARM Sensor difference alarm.
Working temperature and safety sensors
report a temperature difference of more than
35 K.
-07 I2C-BUS ERROR Internal error when reading or writing the I2C
bus.
-08 INVALID COMMAND
Invalid command.
-09 COMMAND NOT ALLOWED IN
CURRENT OPERATING MODE
Invalid command in current operating mode.
-10 VALUE TOO SMALL
Entered value too small.
-11 VALUE TOO LARGE
Entered value too large.
-12 TEMPERATURE MEASUREMENT
ALARM
Error in A/D converter.
-13 WARNING : VALUE EXCEEDS
TEMPERATURE LIMITS
Value lies outside the adjusted range for the
high and low temperature warning limits. But
value is stored.
-14 EXCESS TEMPERATURE
PROTECTOR ALARM
Excess temperature protection alarm
-20 WARNING: CLEAN CONDENSOR OR
CHECK COOLING WATER CIRCUIT OF
REFRIGERATOR
Cooling of the condenser is affected.
Clean air-cooled condenser.
Check the flow rate and cooling water
temperature on water-cooled condenser.
-21 WARNING: COMPRESSOR STAGE 1
DOES NOT WORK
Compressor stage 1 does not work.
-22 WARNING: COMPRESSOR STAGE 2
DOES NOT WORK
Compressor stage 2 does not work.
-23 WARNING: HIGH TEMPERATURE
ON COMPRESSOR STAGE 1
Excess temperature on compressor stage 1.
-24 WARNING: HIGH TEMPERATURE
ON COMPRESSOR STAGE 2
Excess temperature on compressor stage 2.
-25 REFRIGERATOR WARNING
Error in the refrigerating machine.
-30 CONFIGURATION ERROR:
CONFIRM BY PRESSING <ENTER> ON
CIRCULATOR
The configuration of the circulator does not
conform to its present use.
Press to automatically perform a single
modification of the configuration.
-33 SAFETY SENSOR ALARM Excess temperature sensor short-circuited or
interrupted.
-40 NIVEAU LEVEL WARNUNG Low liquid level warning in the internal
reservoir.
47

Table of Contents