HE / SE
9
2.2. Technical specifications
Heating circulators
HE-4 SE-6 SE-12 SE-26
Working temperature range
°C
20 … 250
20 ... 300
20 ... 300
20 ... 300
Temperature stability
°C
±0.01
±0.01
±0.01
±0.01
Bath opening (WxL)
cm
13x15 13x15 22x15 22x30
Bath depth
cm
15
20
20
20
Filling volume
liters
3 ... 4,5 4,5 ... 6 7,5 ... 12 18 ... 26
Overall dimensions (WxDxH)
cm
21x42x40
21x43x44
30x43x47
36x61x47
Weight
kg
11 13,5 14,0 27,0
Bridge mounted circulator
SE-Z
Working temperature range
°C
20 ... 300
Temperature stability
°C
±0.01
Immersion depth max./min.
cm
18/9
Overall dimensions (WxDxH)
cm
32x17x33
Weight
kg
7
HE
SE
Mains power connection
V/Hz
208-230 / 50/60
208-230 / 50/60
Current draw (at 208 V / 230 V)
A
8 / 9
13 / 14
Mains power connection
V/ Hz
100-115 / 50/60
Current draw (at 100 V / 115 V)
A
9 / 10
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C bath fluid: water operating temperature: 70 °C
Technical changes without prior notification reserved.
HE
SE
Temperature selection
digital
via keypad
indication on VFD COMFORT-DISPLAY
remote control via PC
indication on monitor
Temperature indication
VFD COMFORT-DISPLAY
Resolution
°C
0.01
0.01
ATC3 INT/EXT °C ±3 / ±9 ±3 / ±9
Temperature control
ICC - Intelligent Cascade Control
Heater wattage (at 230 V)
kW
2,0
3,0
Heater wattage (at 115 V)
kW
1,0
---
Electronically adj. pump capacity stages
1 ... 4
1 ... 4
Flow rate at 0 bar
l/min
22 ... 26
22 ... 26
Max. pressure at 0 liters
bar
0.7
0.7
Max. suction at 0 liters
bar
0.4
0.4
Electrical connections:
See page 10
Ambient temperature
°C
5 … 40
5 ... 40

Table of Contents