Press SHIFT-DMM SETUP to *ccess modes below. Use
NEXT to scroll through setup menu.
REV--Briefly displays current software revision level.
MUX ON/OFF-Use uprange or downrange to select on/off.
Operating unit in MUX OFF turns off autocal routines, which
speeds up reading rate and minimizes noise spikes at in-
put. Accuracy is reduced in MUX OFF.
IEEE-Use numeric keys to program IEEE-488 primary ad-
dress JO-30). Use SAVE to permanently change address.
FREQ=50/60HZ-Use uprange or downrange to select
operating frequency. Selected frequency must agree with
power line frequency for optimum noise performance.
SAVE YES/NO-Select operating modes then SAVE YES to
store present instwment configuration. Unit will assume
the saved configuration upon power up, or after receiving
IEEE-488 DCL or SDC commands.
LEDS ON/OFF-Use uprange or downrange then press
NEXT to ten front panel LEDS.
DEBUG ON/OFF-Use uprange or downrange then press
NEXT to enter diagnostic program. Press any key except
TRIGGER to exit program.
RESET YES/NO-Select RESET YES with uprange or
downrange to return the unit to factory default conditions.
Use SAVE after RESET to restore power up default to fac-
tory default conditions.
9
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Other manuals for Keithley 199

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 199 and is the answer not in the manual?

Keithley 199 Specifications

General IconGeneral
BrandKeithley
Model199
CategoryScanner
LanguageEnglish