12-14 SCPI Tables
:SENSe2
:PCURrent
:TIME
:DIGitize? Query integration time for digitizing or burst
measurements.
:SYNChronize Path for pulse detection triggering:
[:STATe] <b> Send ON to select pulse current measurements, or
OFF to
select pulse current digitization.
ON
[:STATe]? Query pulse current synchronization selection.
:DELay <NRf> Specify trigger delay in seconds (dependent on
SYNC:ST
AT):
pulse-current measurements: 0 to 0.1 (sync ON)
pulse current digitization: 0 to 5 (sync OFF)
0
:DELay? Query specified trigger delay (dependent on
SYNC:ST
AT).
:TLEVel <NRf> Set trigger level value 0–5A. 0
:TLEVel? Query trigger level setting.
:FAST <b> Enable or disable pulse current fast readings. OFF
:FAST? Query pulse current fast reading setting.
:SEARch <b> Enable or disable pulse current search. ON
:SEARch? Query pulse current search setting.
:DETect <b> Enable or disable pulse current detection mode. OFF
:DETect? Query pulse current detection mode setting.
:TimeOUT <NRf> Specify length of timeout: 5ms - 32s incrementing in
1ms.
1 (sec)
:TimeOUT? Query pulse current timeout setting.
:LINTegration Path to configure long integration measurements (refer to
Section 4):
:TEDGe <name> Set trigger edge to start long integration measurement
(RISIN
G, FALLING, NEITHER).
RISING
:TEDGe? Query trigger edge setting.
:TIME <NRf> Set integration time in seconds: X to 60 where X is 0.850
fo
r 60Hz or 0.840 for 50Hz).
1.00
:TIME? Query integrations time.
Table 12-4 (cont.)
SENSe command summary (refer to Tables 2-3, 3-2, and 4-2) (cont.)
Command Description
Default
parameter
SCPI