Test the instrument
The following test veries basic operation of the 2601B, 2602B,
and 2604B. In this test, you will use the front-panel controls
shown below to source voltage and measure the voltage output.
You do not need to connect a device under test (DUT) for this
test.
Step 1: Set source function, range, and level
1. Press the SRC key. You will see a blinking character in the
SrcA value eld. Conrm that mV is displayed; if not, press the
SRC key again.
2. While that character is still blinking, press the up or down
RANGE keys until 40 V is displayed.
If the instrument has two channels (2602B or 2604B) and
is in dual-channel display mode, press the DISPLAY button
once to show only the settings for SMU channel A.