Models 2601B, 2602B and 2604B
Keithley Instruments, Inc. System SourceMeter
®
28775 Aurora Road
Instrument Specifications
Cleveland, Ohio 44139
1-888-KEITHLEY
http://www.keithley.com
Specifications are subject to change without notice
SPEC-2602B Rev. B / February 2014 1 of 13
SPECIFICATION CONDITIONS
This document contains specifications and supplemental information for the Models 2601B, 2602B and
2604B System SourceMeter
®
instrument. Specifications are the standards against which the Models
2601B, 2602B and 2604B are tested. Upon leaving the factory, the Models 2601B, 2602B and 2604B meet
these specifications. Supplemental and typical values are nonwarranted, apply at 23 °C, and are provided
solely as useful information.
Source and measurement accuracies are specified at the Models 2601B, 2602B and 2604B terminals
under these conditions:
1. 23 °C ± 5 °C, < 70 percent relative humidity
2. After a two-hour warm-up period
3. Speed normal (1 NPLC)
4. A/D autozero enabled
5. Remote sense operation or properly zeroed local operation
6. Calibration period: One year
DC POWER SPECIFICATIONS
Voltage Current
Maximum output
power and
source/sink limits
1
40.4 W maximum
± (40.4 V at 1.0 A, -1.0 A)
± (6.06 V at 3.0 A, -3.0 A)
Four-quadrant source or sink operation
40.4 W maximum
± (1.01 A at 40 V, -40 V)
± (3.03 A at 6 V, -6 V)
Four-quadrant source or sink operation
Refer to the “Pulse Characteristics” section for pulsing details, such as duty cycle and pulse width.
1
Full power source operation regardless of load to 30 °C ambient temperature. Above 30 °C or power sink operation, refer to “Operating
Boundaries” in the Series 2600B Reference Manual for additional power derating information.
Question and Answer IconNeed help?

Do you have a question about the Keithley System SourceMeter 2601B and is the answer not in the manual?

Keithley System SourceMeter 2601B Specifications

General IconGeneral
BrandKeithley
ModelSystem SourceMeter 2601B
CategoryMeasuring Instruments
LanguageEnglish

Summary

Specification Conditions

DC Power Specifications

Voltage Accuracy Specifications

Current Accuracy Specifications

Supplemental Characteristics

Pulse Characteristics

Details specifications for pulse generation, including region limits, pulse width, and duty cycle.

Additional Source Characteristics

Additional Measurement Characteristics

Additional Characteristics

High Capacitance Mode

Measurement Speed Characteristics

Triggering and Synchronization Characteristics

Triggering

Specifies timing parameters for trigger events and their relationship to output changes.

Synchronization

Details synchronization timing for multi-node and single-node source changes.

Supplemental Information

Front-panel Interface

Describes the capabilities of the device's display and keypad for user interaction.

Programming

Covers the embedded scripting engine and user memory availability for control.

Test Script Builder

Overview of the IDE for building, running, and managing test scripts.

TSP Express (embedded)

Details the tool for performing I-V tests without programming.

Software Interface

Lists supported software and environments for instrument interaction.

General Specifications