EasyManuals Logo

Keithley System SourceMeter 2601B User Manual

Keithley System SourceMeter 2601B
13 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #6 background imageLoading...
Page #6 background image
Models 2601B, 2602B and 2604B System SourceMeter
®
Instrument Specifications
Specifications are subject to change without notice
6 of 13 SPEC-2602B Rev. B / February 2014
ADDITIONAL MEASUREMENT CHARACTERISTICS
Contact check
specifications
13,14
Speed
Maximum measurement time
to memory for 60 Hz (50 Hz)
Accuracy (1 year)
23° C ± 5° C
± ( % reading + ohms)
Fast 1.1 ms (1.2 ms) 5 % + 10 Ω
Medium 4.1 ms (5 ms) 5 % + 1 Ω
Slow 36 ms (42 ms) 5 % + 0.3 Ω
Current measure settling
time
15
Time required to reach within 0.1 % of final value after source level command is
processed on a fixed range
Values below for Vout = 1 V
Current range Settling time
1 mA < 100 µs
Input impedance
> 10
ADDITIONAL CHARACTERISTICS
Maximum load impedance Normal mode
10 nF
High-capacitance mode
50 µF
Common mode voltage
250 V DC
Common mode isolation
> 1 GΩ
< 4500 pF
Sense high input impedance
> 10
GΩ
Maximum sense lead
resistance
1 kΩ for rated accuracy
Overrange
101 % of source range
102 % of measure range
13
Includes measurement of SENSE HI to HI and SENSE LO to LO contact resistances.
14
Contact check is not available with the Model 2604B.
15
Compliance equal to 100 mA

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley System SourceMeter 2601B and is the answer not in the manual?

Keithley System SourceMeter 2601B Specifications

General IconGeneral
BrandKeithley
ModelSystem SourceMeter 2601B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals