Models 2601B, 2602B and 2604B
Keithley Instruments, Inc. System SourceMeter
®
28775 Aurora Road
Instrument Specifications
Cleveland, Ohio 44139
1-888-KEITHLEY
http://www.keithley.com
Specifications are subject to change without notice
SPEC-2602B Rev. B / February 2014 1 of 13
SPECIFICATION CONDITIONS
This document contains specifications and supplemental information for the Models 2601B, 2602B and
2604B System SourceMeter
®
instrument. Specifications are the standards against which the Models
2601B, 2602B and 2604B are tested. Upon leaving the factory, the Models 2601B, 2602B and 2604B meet
these specifications. Supplemental and typical values are nonwarranted, apply at 23 °C, and are provided
solely as useful information.
Source and measurement accuracies are specified at the Models 2601B, 2602B and 2604B terminals
under these conditions:
1. 23 °C ± 5 °C, < 70 percent relative humidity
2. After a two-hour warm-up period
3. Speed normal (1 NPLC)
4. A/D autozero enabled
5. Remote sense operation or properly zeroed local operation
6. Calibration period: One year
DC POWER SPECIFICATIONS
Voltage Current
Maximum output
power and
source/sink limits
1
40.4 W maximum
± (40.4 V at 1.0 A, -1.0 A)
± (6.06 V at 3.0 A, -3.0 A)
Four-quadrant source or sink operation
40.4 W maximum
± (1.01 A at 40 V, -40 V)
± (3.03 A at 6 V, -6 V)
Four-quadrant source or sink operation
Refer to the “Pulse Characteristics” section for pulsing details, such as duty cycle and pulse width.
1
Full power source operation regardless of load to 30 °C ambient temperature. Above 30 °C or power sink operation, refer to “Operating
Boundaries” in the Series 2600B Reference Manual for additional power derating information.
Question and Answer IconNeed help?

Do you have a question about the Keithley System SourceMeter 2602B and is the answer not in the manual?

Keithley System SourceMeter 2602B Specifications

General IconGeneral
Model2602B
TypeSourceMeter
Channels2
Voltage - Max40 V
Current - Max Source3 A
Measurement Accuracy0.015%
Resolution6.5 digits
Communication InterfaceGPIB, USB, Ethernet
Current Range10nA to 3A
Voltage Resolution100nV
DisplayLCD

Summary

Specification Conditions

DC Power Specifications

Voltage Accuracy Specifications

Current Accuracy Specifications

Supplemental Characteristics

Pulse Characteristics

Minimum Programmable Pulse Width

Additional Source Characteristics

Voltage Output Headroom

Over-temperature Protection

Limit Compliance

Voltage Source Output Settling Time

Current Source Output Settling Time

Additional Measurement Characteristics

Additional Characteristics

High Capacitance Mode

Measurement Speed Characteristics

Triggering and Synchronization Characteristics

Supplemental Information

Reading Buffers

System Expansion

Timing

General Specifications

Digital IO Interface