2600S-901-01 Rev. C / January 2008 Return to Section Topics 5-11
Series 2600 System SourceMeter® Instruments Reference Manual Section 5: Sweep Operation
Custom sweep example
1. Configure source functions
Examples – The following commands restore defaults and set the compliance to 10mA:
smua.reset() -- Restore Series 2600 defaults.
smua.source.limiti = 10e-3 -- Set compliance to 10mA.
2. Configure and execute the sweep.
Example – The following parameters configure a list sweep with 3V, 1V, 4V, 5V, 2V points
using a 0.1s settling time:
vlist = {3, 1, 4, 5, 2}
-- Define voltage list.
SweepVListMeasureI(smua, vlist, 0.1, 5)
-- List sweep, channel A, 3V, 1V, 4V, 5V, 2V steps, 0.1s delay, 5 points.
waitcomplete()
-- Wait for sweep to complete.
3. Request readings.
Request readings from Buffer 1 as follows:
printbuffer(1, 5, smua.nvbuffer1.readings)