2600S-901-01 Rev. C / January 2008 Return to Section Topics 10-19
Series 2600 System SourceMeter® Instruments Reference Manual Section 10: Digital I/O and Triggering
Either Edge trigger mode
The Either Edge trigger mode generates a low pulse and detects both rising and falling edges.
Figure 10-11: Either Edge input trigger
Input characteristics:
All rising or falling edges generate an input trigger event
Figure 10-12: Either Edge output trigger
Output characteristics:
•The trigger.assert command generates a low pulse that is similar to the Falling Edge
trigger mode