2600S-901-01 Rev. C / January 2008 Return to Section Topics 13-33
Series 2600 System SourceMeter® Instruments Reference Manual Section 13: Factory Scripts
Example(s)
smua.reset()
smua.source.rangev = 5
smua.source.rangei = 1
smua.source.levelv = 0
smua.measure.rangev = 5
smua.measure.rangei = 1
smua.measure.nplc = 0.01
smua.measure.autozero = smua.AUTOZERO_ONCE
smua.nvbuffer1.clear()
smua.nvbuffer1.appendmode = 1
smua.source.output = smua.OUTPUT_ON
smub.reset()
smub.source.func = smub.OUTPUT_DCAMPS
smub.source.rangei = 1
smub.source.rangev = 5
smub.source.leveli = 0
smub.measure.rangei = 1
smub.measure.rangev = 5
smub.measure.nplc = 0.01
smub.measure.autozero = smub.AUTOZERO_ONCE
smub.nvbuffer1.clear()
smub.nvbuffer1.appendmode = 1
smub.source.output = smub.OUTPUT_ON
Table 13-19 (continued)
KIGeneral TSP test script: InitiatePulseTestDual
TSP project name: KIFactoryPulse
TSP test script name: KIPulse
Firmware version: 1.2.0 and later
Function:
InitiatePulseTestDual
Usage:
f, msg = InitiatePulseTestDual(tag1, tag2)