EasyManuals Logo

Keithley 2635 Reference Manual

Keithley 2635
562 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #429 background imageLoading...
Page #429 background image
2600S-901-01 Rev. C / January 2008 Return to Section Topics 14-15
Series 2600 System SourceMeter® Instruments Reference Manual Section 14: Display Operations
display.waitkey
The display.waitkey function captures the keycode value for the next key press:
key = display.waitkey()
After sending the display.waitkey function, the script will pause and wait for the operator to
press a front panel key. For example, if the MEAS key is pressed, the function will return the value
86, which is the keycode for that key. The keycode values are listed in
Table 14-3.
Example: The following code will prompt the user to press the EXIT key to abort the script, or any
other key to continue it:
display.clear()
display.setcursor(1, 1)
display.settext("Press EXIT to Abort")
display.setcursor(2, 1)
display.settext("or any key to continue")
key = display.waitkey()
display.clear()
display.setcursor(1, 1)
if (key == 75) then
display.settext("Test Aborted")
exit()
else display.settext("Test Continuing")
end
The above code captures the key that is pressed by the operator. The keycode value for the EXIT
key is 75. If EXIT is pressed, the script aborts. If any other key is pressed, the script will continue.

Table of Contents

Other manuals for Keithley 2635

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2635 and is the answer not in the manual?

Keithley 2635 Specifications

General IconGeneral
BrandKeithley
Model2635
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals