EasyManua.ls Logo

Keithley 2635

Keithley 2635
562 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
www.keithley.com
Specifications are subject to change without notice.
SPEC-2601_2602 Rev. D / January 2008 Page 3 of 7
Models 2601/2602
System SourceMeter
®
Specifications
Additional Source Specifications
Transient response time: <70Čťs for the output to recover to 0.1% for a 10% to 90% step change in load.
Voltage source output settling time: Time required to reach 0.1% of final value, when changing from 10% to 90%
of range, after source level command is processed on a fixed range.
100mV, 1V ranges: <50Čťs typical.
6V Range: <100Čťs typical.
40V Range: <150Čťs typical.
5
Current source output settling time: Time required to reach 0.1% of final value, when changing from 10% to 90%
of range, after source level command is processed on a fixed range. Values below for Iout · Rload = 1V
unless noted.
3A–10mA ranges: <80ȝs typical (current less than 2.5A, Rload >1.5ȍ).
1mA range: <100Čťs typical.
100ČťA range: <150Čťs typical.
10ČťA range: <500Čťs typical.
1ČťA range: <2.5ms typical.
100nA range: <25ms typical.
DC floating voltage: Output can be floated up to ±250VDC from chassis ground.
Remote sense operating range:
1
Maximum voltage between HI and SENSE HI = 3V.
Maximum voltage between LO and SENSE LO = 3V.
Voltage output headroom:
40V range: Max. output voltage = 42V – total voltage drop across source leads (maximum 1ȍ per source lead).
6V range: Max. output voltage = 8V – total voltage drop across source leads.
Over temperature protection: Internally sensed temperature overload puts unit in standby mode.
Voltage source range change overshoot: Overshoot into a 100kČŤ load, 20MHz BW, 300mV typical.
Current source range change overshoot: <5% of larger range + 300mV/Rload typical.
(see “CURRENT SOURCE OUTPUT SETTLING TIME” for additional test conditions).
5 Add 150Čťs when measuring on the 1A range.

Table of Contents

Other manuals for Keithley 2635

Related product manuals