4904 SERIOUS ICX parameter count mismatch, %s (Line #%d)
4905 SERIOUS ICX parameter invalid value, %s (Line #%d)
4906 SERIOUS ICX invalid function id, %s (Line #%d)
5001 FATAL SMU is unresponsive
5003 SERIOUS Saved calibration constants corrupted
5004 SERIOUS Operation conflicts with CALA sense mode
5005 SERIOUS Value too big for range
5007 SERIOUS Operation would exceed safe operating area of the instrument
5008 SERIOUS Operation not permitted while output is on
5009 SERIOUS Unknown sourcing function
5010 SERIOUS No such SMU function
5011 SERIOUS Operation not permitted while cal is locked
5012 SERIOUS Cal data not saved - save or restore before lock
5013 SERIOUS Cannot save cal data - unlock before save
5014 SERIOUS Cannot restore cal data - unlock before restore
5015 SERIOUS Save to cal set disallowed
5016 SERIOUS Cannot change cal date - unlock before operation
5017 SERIOUS Cannot change cal constants - unlock before operation
5018 SERIOUS Cal version inconsistency
5019 SERIOUS Cannot unlock - invalid password
5021 SERIOUS Cannot restore default calset. Using previous calset
5022 SERIOUS Cannot restore previous calset. Using factory calset
5023 SERIOUS Cannot restore factory calset. Using nominal calset
5024 SERIOUS Cannot restore nominal calset. Using firmware defaults
5025 SERIOUS Cannot set filtercount > 1 when measure.count > 1
5027 SERIOUS Unlock cal data with factory password
5028 SERIOUS Cannot perform requested operation while source autorange is
enabled
5029 SERIOUS Cannot save without changing cal date and cal due values
5032 RECOVERABLE Cannot change this setting unless buffer is cleared
5033 RECOVERABLE Reading buffer not found within device
5038 RECOVERABLE Index exceeds maximum reading
5040 RECOVERABLE Cannot use same reading buffer for multiple overlapped
measurements
5041 SERIOUS Output Enable not asserted
5042 SERIOUS Invalid while overlapped measure
5043 SERIOUS Cannot perform requested operation while voltage measure
autorange is enabled
5044 SERIOUS Cannot perform requested operation while current measure
autorange is enabled
5045 SERIOUS Cannot perform requested operation while filter is enabled
5046 SERIOUS SMU too hot
5047 RECOVERABLE Minimum timestamp resolution is 1µs
5048 SERIOUS Contact check not valid with HIGH-Z output off
5049 SERIOUS Contact check not valid while an active current source
5050 SERIOUS I limit too low for contact check
5051 FATAL Model Number/SMU Hardware mismatch
5052 RECOVERABLE Interlock engaged; system stabilizing
5052 RECOVERABLE Cannot disable outputenableaction
Table B-2 (cont.)
Error summary
Error number Error level Error Message