Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-44 3700AS-901-01 Rev. D/June 2018
channel.calibration.unlock()
This function unlocks calibration functionality for a card so that calibration operations can be performed.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
channel.calibration.unlock("slotX", password)
X
The slot number ( 1 to 6)
password
The password that unlocks calibration
Details
Calibration data is locked during normal operation. This command enables calibration functionality.
When calibration is completed, calibration functionality must once again be locked
(channel.calibration.lock()). Only one card at a time may be unlocked.
There is only one password per card. Therefore, the only acceptable values for channel list are
"slot1", "slot2", and so on. Otherwise, an error is generated.
An error is generated if the password that is entered does not match the one that was saved with
channel.calibration.password().
The password can only contain six case-sensitive, alphanumeric characters.
The default password from the factory is KI3706.
Example
channel.calibration.unlock("slot1","KI3706")
-- Perform operations to generate the calibration data
channel.calibration.save()
channel.calibration.lock()
Unlock the channel.
Save the calibration settings.
Lock the channel.
Also see
channel.calibration.lock() (on page 11-40)
channel.calibration.password() (on page 11-40)
channel.calibration.save() (on page 11-41)