The default powerline frequency is set to 60 Hz. If the setting
is wrong, the 4200A-SCS cannot properly reject powerline
measurement noise.
To change the powerline frequency:
1. Close Clarius.
2. Run KCon.
3. Select KI 4200A SCS in the Con guration Navigator. The left
pane is the Con guration Navigator.
4. Change the Powerline Frequency as needed.
5. Select Save .
Change powerline frequency