OPERATION
1. Select function and range.
2. Input the offset or baseline level to the Model 617.
3. Press the SUPPRESS button. The SUPPRESS indicator light will
turn on and the display will zero.
4. The stored level will be algebraically subtracted from all subsequent
measurements.
NOTE: Suppress can be turned off by again pressing the SUPPRESS
button.
TYPICAL USE
1. Cancel external source offset (i.e. calibrator offset).
2. Null out lead resistance on ohms.
3. Null out circuit outputs to display only drift or changes.