DATA STORE with MIWMAX
DATA STORE
RECALL EXIT
DESCRIPTION
The data store feature saves up to 100 points of data and holds the
lowest (Anin) and highest (Max) readings while data store is on. The 100
points of data are stored at the rate of one of six selectable rates from
every reading (3 per second) to one reading per hour. Also, manual data
entry is available (RBI.. In the continuous trigger mode (SGL light off),
the minimumlmaximum readings are continuously updated at the rate
of three readings per second (regardless of the selected rate) while data
store is on.
100 POINT DATA STORAGE
1. Press and hold in the Data Store ON/OFF button. The following
storage rates will scroll on the display:
r=O (every reading)
r= 1 (1 reading per second)
r=2 (1 reading per 10 seconds)
r=3 (1 reading per minute)
r=4 (1 reading per 10 minutes)
r=5 (1 reading per hour)
r =6 (every time TRIG button is pressed)
2. When the desired storage rate is displayed, release the Data Store
ON/OFF button.
23
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com