HP 9816
The following program sends a command string to the Model 617 Elec-
trometer, reads data and displays the data on the HP 9816 CRT, using
BASICA 2.0.
DIRECTIONS
1. Using the front panel pmgram feature, set the primary address of the
Model 617 to 27.
2. With the power off, connect the Model 617 to the HP 9816 and HP
82937A GPIB interface,
3. Type EDlT and press the EXEC key.
4. Enter the following program using the ENTER key after each line is
typed.
5. Press the HP 9816 RUN key.
6. The display will read “TEST SETUP”.
7. To program the Model 617 to the w range and to take a reading,
type in FlR7TlX and press the ENTER key.
8. The display will read NDCA + O.OOOOOE-06 when the Model 617 is
in zero check.
PROGRAM
COMMENTS
10 REMOTE 727
Set to remote.
20 INPUT’ITEST SETuP”,A$
Prompt for test setup.
30 OUPTUT 727;AS
Send command string to 617.
40 ENTER 727;B$
Get data string from 617.
50 PRINT BS
Display data string.
64 GOT020 Repeat.
70 END
NOTE: If conversion to numeric variable is desired, change lines 40 and
50 as follows:
40 ENTER 727;B
50 PRINT B