VOLTAGE SOURCE ACCURACY TEMPERATURE
(1 Year)
1
COEFFICIENT
5½
DIGIT 18°–28°C0°–18°C & 28°–50°C
RANGE RESOLUTION ±(% setting + offset) ±(% setting+offset)/°C
100 V 5 mV 0.15 + 10 mV 0.005 + 1 mV
1000 V 50 mV 0.15 + 100 mV 0.005 +10 mV
MAXIMUM OUTPUT CURRENT:
±10mA; active current limit at <11.5mA for 100V range.
±1mA; active current limit at <1.15mA for 1000V range.
SETTLING TIME:
<8ms to rated accuracy for 100V range.
<50ms to rated accuracy for 1000V range.
NOISE:
<150µV p-p from 0.1Hz to 10Hz for 100V range.
<1.5mV p-p from 0.1Hz to 10Hz for 1000V range.
TEMPERATURE (THERMOCOUPLE)
ACCURACY (1 Year)
1
THERMOCOUPLE 18°–28°C
TYPE RANGE ±(% rdg + °C)
K –25°C to 150°C ± (0.3% ± 1.5°C)
Note:
1
Excluding probe errors, T
CAL
± 5°C, 1 PLC integration time.
HUMIDITY ACCURACY (1 Year)
1
18°–28°C
RANGE ±(% rdg + % RH)
0–100% ± (0.3% +0.5)
Note:
1
Humidity probe accuracy must be added. This is ±3% RH, for Model 6517-RH, up to 65°C probe environment, not to exceed 85°C.
IEEE-488 BUS IMPLEMENTATION
MULTILINE COMMANDS: DCL, LLO, SDC, GET, GTL, UNT, UNL, SPE, SPD.
IMPLEMENTATION: SCPI (IEEE-488.2, SCPI-1994.0); DDC (IEEE-488.1).
UNILINE COMMANDS: IFC, REN, EOI, SRQ, ATN.
INTERFACE FUNCTIONS: SH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E1.
PROGRAMMABLE PARAMETERS: Function, Range, Zero Check, Zero Suppress, EOI (DDC mode only), Trigger, Terminator
(DDC mode only), 100-Reading Storage (DDC mode), 15706 Max. Reading Storage (SCPI mode), Calibration (SCPI mode
only), V-Source Output, Display Format, SRQ, Status (including V-Source I-Limit), Output Format, Guard.
ADDRESS MODES: TALK ONLY and ADDRESSABLE.
TRIGGER TO READING DONE: 150ms typical, with external trigger.
RS-232 IMPLEMENTATION:
Supports: SCPI 1994.0.
Baud Rates: 300, 600, 1200, 2400, 4800, 9600, 19.2k.
Protocols: Xon/Xoff, 7 or 8 bit ASCII, parity-odd/even/none.
Connector: DB-9 TXD/RXD/GND.
4-4 Specifications