2400 Series SourceMeter
®
User’s Manual Basic Source-Measure Operation 3-5
2420 2425/2430
Range Source Measure Range Source Measure
200mV
2V
20V
60V
10µA
100µA
1mA
10mA
100mA
1A
3A
±210mV
±2.1V
±21V
±63V
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
±3.15A
±211mV
±2.11V
±21.1V
±63.3V
±10.55µA
±105.5µA
±1.055mA
±10.55mA
±105.5mA
±1.055A
±3.165A
200mV
2V
20V
100V
10µA
100µA
1mA
10mA
100mA
1A
3A/10A
±210mV
±2.1V
±21V
±105V
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
*
±211mV
±2.11V
±21.1V
±105.5V
±10.55µA
±105.5µA
±1.055mA
±10.55mA
±105.5mA
±1.055A
**
Max Power = 66W 2425 and 2430 Max Power
DC Mode: 110W (105V, 1.05A)
66W (21V, 3.15A)
Pulse Mode: 1.1kW (2430 only)
2440
Range Source Measure
200mV
2V
10V
40V
10µA
100µA
1mA
10mA
100mA
1A
5A
±210mV
±2.1V
±10.5V
±42V
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
±5.25A
±211mV
±2.11V
±10.5V
±42V
±10.55µA
±105.5µA
±1.055mA
±10.55mA
±105.5mA
±1.055A
±5.25A
Max Power = 66W
* ±3.15A (DC mode; 2425 and 2430)
±10.5A (Pulse mode; 2430 only)
** ±3.165A (DC mode; 2425 and 2430)
±10.55A (Pulse mode; 2430 only)
Table 3-1 (continued)
Source-measure capabilities
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2400

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 and is the answer not in the manual?

Keithley SourceMeter 2400 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals