EasyManuals Logo

KERN ALJ Operating Instruction

KERN ALJ
Go to English
304 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #114 background imageLoading...
Page #114 background image
TALJG_A/TALSG_A/TPLJG_A/TPLSG_A-BA-e-2112 11
KERN PLJ 720-3AM PLJ 6200-2AM
Product number / type TPLJG 720-3M-A TPLJG 6200-2M-A
Weighing range (Max) 720 g 6200 g
Interval (d) 0.001 g 0.01 g
Reproducibility 0.001 g 0.01 g
Linearity ±0.002 g ±0,02 g
Verification scale interval (e) 10 mg 100 mg
Verification class II II
Minimum weight (Min) 20 mg 50 mg
Settling time (standard) 2 s 2 s
Minimum part weight when
counting the number of pieces in
laboratory conditions**
1 mg 10 mg
Minimum part weight when
counting the number of pieces in
standard conditions**
10 mg 100 mg
Heating time 4 h 4 h
Adjustment weight internal
Number of reference pieces
when counting the number of
pieces
10, 25, 50, 100, selected freely
Weight units ct, g
Power supply
230V/50Hz AC (Euro), 24V/1A DC
Operating temperature +15°C ... +30°C
Air humidity max. 80% (non-condensing)
Housing (W × D × H) mm
210 x 345 x 155 210 x 345 x 160
Wind breaker [mm]
internal: Ø 150, height 60
external: Ø 160, height 70
Scale plate (stainless steel) Ø 11 cm Ø 16 cm
(Net) weight kg
4.9 kg 5.4 kg
Interface RS-232C
Contamination degree 2
Overvoltage category Category II
Installation height
above sea level
up to 4000 m
Location site Solely indoors

Table of Contents

Other manuals for KERN ALJ

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN ALJ and is the answer not in the manual?

KERN ALJ Specifications

General IconGeneral
BrandKERN
ModelALJ
CategoryScales
LanguageEnglish

Related product manuals