19 FFN-N-BA-e-1323
7.1.1 Adjusting non-verifiable models
Display
Operation
Start balance by pressing
Whilst the weighing scale is carrying out a self-test,
(00…- 99…) press and hold until the message “F1
CAL“ appears.
Press ,
in the display appears „UnLod“.
(Example)
Press again
the weight value will be displayed.
Enter value for adjustment weight (See chpt 1) by
Position weight
Press
Remove weight whilst scale is carrying out self-test
Weighing scale changes to zero display.
This completes the adjustment process.