FKB_A-BA-e-1615
29
Animal
weighing
function
0.0
g
Vn�
T
ANL
The animal weighing function can be applied for busy weighing.
During a defined period the average value of the weighing results
is formed.
The more unquiet the weighed item, the longer the
period should
be selected.
In weighing mode keep the PRINT key pressed until [Unit]
appears.
Press the MODE button several times until „ANL“ is displayed.
Acknowledge using SET key, the current setting is displayed.
Select the desired settings by pressing the MODE key
ANL 3
Period 3 seconds
ANL 5
Period 5 seconds
ANL 10
Period 10 seconds
ANL 15
Period 15 seconds
ANL off
Animal weighing not active
Use the SET key to confirm selection.
Put the weighing good (animal) on the weighing plate and press
the SET button. In the display runs a „Countdown“.
The average value of the weighing results is displayed and
remains displayed on the screen.
Use the SET key to change between animal weighing and
normal weighing.
Press the SET key to restart the weighing cycle for animal
weighing.

Other manuals for KERN FKB A Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN FKB A Series and is the answer not in the manual?

KERN FKB A Series Specifications

General IconGeneral
BrandKERN
ModelFKB A Series
CategoryScales
LanguageEnglish

Related product manuals