Module 2
Basic Measurement
2-10
Tracer Test is the best way for performing I-V measurement easily and quickly.
Display the Tracer Test screen and set the measurement condition.
You can finish the setup by the following five steps.
1. Click Add button and add SMU1, SMU2, SMU3, and SMU4.
2. Set the channel definition. See below.
3. Set the X axis parameter, polarity and maximum value.
4. Set the Y axis parameter, polarity and maximum value.
5. Define the VAR1, VAR2, and CONST setup. See below.
Channel definition:
SMU1: Vsub, Isub, V mode, CONST
SMU2: Vs, Is, V mode, CONST
SMU3: Vg, Ig, V mode, VAR2
SMU4: Vd, Id, V mode, VAR1
Source setup:
VAR1: Stop 0 V, Compliance 100 mA
VAR2: Start 2 V, Stop 5 V, NOS 4, Compliance 10 mA
CONST: Compliance 100 mA