2 Calibration Procedures
46 Keysight DAQ970A/DAQ973A Service Guide
Calibration Actions
Entering calibration values
Use measurement commands to set the instrument's measurement parameters
and functions for the mode being calibrated, then enter calibration values with the
following:
Front panel operation:
[Utility] > Calibrate > Cal Value
After specifying the input value for the calibration step to be performed, press
Perform Cal Step.
Remote interface operation:
Enter cal value: CALibration:VALue <value>
Perform calibration: CALibration[:ALL]?
Storing calibration constants
To store calibration values in non-volatile memory, use the following:
Front panel operation: [Utility] > Calibrate > Save Cal Values
Remote interface operation: CALibration:STORe