2 Verification and Adjustments
76 Keysight E36200 Series Service Guide
Table 2-12 Test record form - Keysight E36232A
E36232A Report number _______________ Date _______________
Description DUT setting Lower limit Result Upper limit
Constant voltage tests
Voltage programming
Zero voltage output (V
0
)
0 V, 10 A –0.008 V ____________ 0.008 V
Maximum voltage output (V
max
)
60 V 10 A 59.974 V ____________ 60.026 V
Voltage readback
Zero voltage measured over interface 0 V, 10 A
V
0
– 0.008 V
____________
V
0
+ 0.008 V
Maximum voltage measured over interface 60 V 10 A
V
max
– 0.026 V
____________
V
max
+ 0.026 V
CV load regulation (V
load
- V
noload
)
60 V, Max A –8 mV ____________ 8 mV
20 V, Max A –4 mV ____________ 4 mV
CV line regulation (V
lowline
- V
highline
)
60 V, Max A –8 mV ____________ 8 mV
20 V, Max A –4 mV ____________ 4 mV
CV ripple and noise
peak-to-peak
60 V, Max A - ____________ 3.5 mV
20 V, Max A - ____________ 3.5 mV
rms
60 V, Max A - ____________ 0.35 mV
20 V, Max A - ____________ 0.35 mV
Transient response
60 V, Max A - ____________ 50 μs
20 V, Max A - ____________ 50 μs

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E36200 Series and is the answer not in the manual?

Keysight Technologies E36200 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE36200 Series
CategoryPower Supply
LanguageEnglish

Related product manuals