6 Troubleshooting
128 Keysight E4416A/E4417A Power Meters Service Guide
Introduction
This chapter enables qualified service personnel to diagnose suspected faults
with the power meter Rmt I/O (Remote Input/Output) signal lines and RS232/422
serial port.
If there is a problem when attempting to use the RS232/422 serial interface or the
remote I/O functions, consult the User's Guide and confirm that all the user
setups are correct before proceeding with the following fault finding flow charts.
Suggested Diagnostic Equipment
1 Digital multi-meter capable of measuring voltage and resistance, for example,
Keysight E4975.
2 Signal source capable of producing 300 ms TTL single shot pulses, for
example, Keysight 33120A.
3 +5 V TTL logic level source.
4 RS232 self test connector 9 way 'D' type female connector with the following
pins wired together:
Pin 2 (Rx)to Pin 3 (Tx).
Pin 4 (DTR)to Pin 6 (DSR).
Pin 7 (RTS)to Pin 8 (CTS).
5 RS422 self test connector - 9 way 'D' type female connector with the following
pins wired together:
–Pin 1 (CTS)to Pin 9 (RTS).
Pin2 (Rx)to Pin 4 (Tx).
Pin 3 (Tx+)to Pin6 (Rx+).
Pin7 (RTS+)to Pin 8 (CTS+).

Table of Contents

Other manuals for Keysight Technologies EPM Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies EPM Series and is the answer not in the manual?

Keysight Technologies EPM Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelEPM Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals