2-32 Keysight M937xA PXIe Service Guide
Tests, Adjustments, and Troubleshooting
Performance Tests
2-
If the Analyzer Fails this Test
-Perform the “Source Adjustment” on page 2-46 and repeat this test.
- If the analyzer still fails this test, return it to Keysight for repair. Refer to
“Shipping Your Network Analyzer for Service or Repair” on page 1-3.