Keysight M9537A AXIe Embedded Controller Startup Guide 25
Environmental Operating Conditions
Environmental Operating Conditions
The Keysight M9537A AXIe Embedded Controller is designed to operate in a
temperature range of 0° C to +55° C with non-condensing humidity. The
maximum humidity is 95% at +40° C. The maximum operating altitude is 3000 m
(9800 ft). The module should be operated in an indoor environment where
temperature and humidity are controlled. Condensation can pose a potential
shock hazard.
Condensation can occur when you move the module from a cold to a warm
environment, or if the temperature and/or humidity of the environment change
quickly.
Samples of this product have been type tested in accordance with the Keysight
Environmental Test Manual and verified to be robust against the environmental
stresses of storage, transportation and end-use. The stresses included are not
limited to temperature, humidity, shock, vibration, altitude and power line
conditions. For complete specifications, check the Keysight web site at:
www. keysight.com/find/M9537A.