416 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Using BERT for N5172B/82B with Option UN7
Verifying BERT Operation
Measurement Example Using Custom Digital Modulation
(Requires Option 431)
The following steps set up the signal generator for a BERT measurement using Custom Digital
Modulation.
1. Refer to Figure 16-16 and make the following connections on the signal generator’s rear panel.
DATA OUT (AUX I/O connector pin 33) to BER DATA IN (BNC connector labeled EVENT 1).
DATA CLK OUT (AUX I/O connector pin 7) to BER CLK IN (BNC connector labeled BB
TRIG1).
2. Press the Preset key. This configures the signal generator to a pre-defined state.
3. Press the Mode key.
4. Press Real-Time Custom > Modulation Setup > Modulation Type > QPSK.
5. Press Return > Return > Return > Data > PN Sequence > More > PN9.
6. Press Return > More > More > Data/Clk/Sync Signal Polarity Setup > Data/Clk/Sync Rear
Outputs On.
7. Press the Aux Fctn key.
8. Press Digital Bus Connectivity PXB/BERT N5102A to N5102A > Confirm Changes > BERT >
BERT Off On to On > BER Display > BER Display% Exp to% > Display Update Cycle End Cont
to Cont.
The following steps configure BERT measurement parameters.
9. Press Return > Configure BERT > Total Bits to 100000 > Enter > More > Special Pattern Ignore
Off On to Off.
10.Press BERT Resync Off On to Off > Return > Pass/Fail Off On to Off.
11.Press Return > I/O Setup > Gate Control > Gate Off On to Off.
12.Press Return > Polarity Setup > Clock Polarity Pos Neg to Neg > Data Polarity Pos Neg to Pos.
13.Press Return > Return > Configure Trigger > Cycle Count 0 > Enter.
14.
Press Bit Delay Off On to Off > BERT Trigger to Trigger Key.
15.Press the Trigger key. Figure 16-18 shows the signal generator’s front-panel display after
completion of the these steps.

Table of Contents

Other manuals for Keysight Technologies N5172B EXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5172B EXG and is the answer not in the manual?

Keysight Technologies N5172B EXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5172B EXG
CategoryPortable Generator
LanguageEnglish

Related product manuals