Note that you have the ability to change the identity of the mainframe. This functionality is only
intended for compatibility with previous "A"and "B" version mainframes.
Front Panel Menu Reference SCPI Command
Select System\Preferences\IDN
Check the box to identify the instrument as an Agilent
product.
SYST:PERS:MAN "<manufacturer>"
SYST:PERS:MOD "<model number>"
The IDNchange and PERSona commands affects the following identification items:
l *IDN? command for vendor and model
l VISA programmatic access APIs for vendor and model
l LXI instrument web pages
l LXI XML
l LXI mDNS announcements
Instrument State Storage
The power system has ten storage locations in non-volatile memory to store instrument states. The
locations are numbered 0 through 9. Any state previously stored in the same location will be
overwritten.
Front Panel Menu Reference SCPI Command
Select States\SaveRecall.
In the SaveRecall field, enter a location from 0 to 9. Then
press Select.
Select Save to save the state or Recall to recall a state.
To save a state in location 1: *SAV 1
To recall a state from location 1: *RCL 1
Specifying a power-on state
When shipped from the factory, the power system is configured to automatically recall the reset
(*RST) settings at power-on. However, you can configure the power system to recall the settings you
have stored in memory location 0 (RCL0) at power-on.
Front Panel Menu Reference SCPI Command
Select States\PowerOn.
Select Recall State0. Then press Select.
OUTP:PON:STAT RCL0
Output Groups
Output grouping does not apply to Keysight Models N678xA SMU.
4 Operating the Power System
158 Keysight N6700C Series User's Guide