Safety
8
If excess current is input, generated heat activates a built-in 
safety function that blocks normal output. If this happens, 
remove the input immediately (unclamp the sensor from the 
conductor being measured or reduce the input current to zero). 
Wait until the sensor has had sufficient time to cool before 
resuming operation.
Even if the input current does not exceed the rated continuous 
maximum, continuous input for an extended period of time may 
result in activation of the safety circuit to prevent damage 
resulting from heating of the sensor.
At high ambient temperatures, the built-in safety circuit may 
activate at current input levels below the rated continuous 
maximum.
Continuous input of current exceeding the rated maximum or 
repeated activation of the safety function may result in damage 
to the unit.
Do NOT measure current such that the total probe current 
consumption exceeds the allowable AutoProbe interface current 
consumption. The excess current consumption causes a 
temporary shutdown of the Infiniium power supply for safety. 
Quit the measurement and cycle the power of the Infiniium, if the 
shutdown occurs. The typical probe current consumption from 
the AutoProbe interface is shown in Figure 1 on page 9.