5-22 Electronic Calibration Modules Reference Guide
Specifications and Characteristics
Typical Pin Depth Values
5-
Table 5-32 4.3-10 Connector Pin Depth Characteristics
Device Typical Pin Depth
in Millimeters
(10
–4
inches)
Measurement Uncertainty
a
in Millimeters
(10
–4
inches)
Observed Pin Depth Limits
in Millimeters
(10
–4
inches)
4.3-10 mm ECal Module
(N4431D)
–0.0000 to –0.1000
(0.0 to –39.0)
+0.0041 to –0.0041
(+1.6 to –1.6)
+0.0041 to –0.1041
(+1.6 to –40.6)
a. Approximately +2 sigma to –2 sigma of gage uncertainty based on studies performed at the factory using precision digital
gages according to recommended procedures.
Table 5-33 3.5 mm Connector Pin Depth Characteristics
Device
Typical Pin Depth
a
in Millimeters
(10
–4
inches)
Measurement Uncertainty
b
in Millimeters
(10
–4
inches)
Observed Pin Depth Limits
in Millimeters
(10
–4
inches)
3.5 mm ECal Module
(N755xA)
–0.0000 to –0.1270
(–0.0 to –50.0)
+0.0041 to –0.0041
(+1.6 to –1.6)
+0.0041 to –0.1311
(+1.6 to –51.6)
3.5 mm ECal Module
(N4431D, N4433D)
–0.0254 to –0.0508
(–10.0 to –20.0)
+0.0041 to –0.0041
(+1.6 to –1.6)
–0.0213 to –0.0549
(–8.4 to –21.6)
3.5 mm ECal Module
(N4691D)
–0.0025 to –0.0152
(–1.0 to –6.0)
+0.0041 to –0.0041
(+1.6 to –1.6)
+0.0016 to –0.0193
(+0.6 to –7.6)
3.5 mm Adapter 0 to –0.0254
(0 to –10)
+0.0041 to –0.0041
(+1.6 to –1.6)
+0.0041 to –0.0295
(+1.6 to –11.6)
a. On N755xA ECals, the 3.5 mm connector is not a precision contact per IEEE precision connector standards.
b. Approximately +2 sigma to –2 sigma of gage uncertainty based on studies performed at the factory using precision digital
gages according to recommended procedures.
Table 5-34 2.92 mm Connector Pin Depth Characteristics
Device Typical Pin Depth
in Millimeters
(10
–4
inches)
Measurement Uncertainty
a
in Millimeters
(10–4
inches)
Observed Pin Depth Limits
in Millimeters
(10
–4
inches)
2.92 mm ECal Module (N4692D) –0.0025 to –0.0152
(–1.0 to –6.0)
+0.0030 to –0.0030
(+1.2 to –1.2)
+0.0005 to –0.0182
(+0.2 to –7.2)
2.92 mm Adapter 0 to –0.0381
(0 to –15)
+0.0030 to –0.0030
(+1.2 to –1.2)
+0.0030 to –0.0411
(+1.2 to –16.2)
a. Approximately +2 sigma to –2 sigma of gage uncertainty based on studies performed at the factory using precision digital
gages according to recommended procedures.

Table of Contents